Flo- Gard 6301 #55221 11/03/11 12:55 AM | |
Joined: Apr 2005 Posts: 259 Portugal Mofe Master | |
OP Mofe Master Joined: Apr 2005 Posts: 259 Portugal | Hi, I need some help. I Have to repair an infusion pump Baxter Flo-Gard 6301 but appears in display COMMON FAILURE F94, someone can help me? Best regards, |
Re: Flo- Gard 6301 Mofe #55222 11/03/11 2:38 AM | |
Joined: Apr 2009 Posts: 81 hong kong dennis_yim Adept | |
dennis_yim Adept Joined: Apr 2009 Posts: 81 hong kong | this is seem a common fault on 6201. Last edited by dennis_yim; 11/03/11 2:39 AM. |
Re: Flo- Gard 6301 Mofe #55226 11/03/11 6:50 AM | |
Joined: Feb 2004 Posts: 14,483 Likes: 37 the path less trodden
Geoff Hannis Super Hero | |
Geoff Hannis Super Hero
Joined: Feb 2004 Posts: 14,483 Likes: 37 the path less trodden | Check-out this thread, Mofe. We've been there before. |
Moderated by DaveC in Oz, RoJo
Hop To
Link Copied to Clipboard
Who's Online Now |
2 members (Huw, fanie), 104 guests, and 178 robots. |
Key: Admin, Global Mod, Mod |
Newest Members |
JonnyP83, carl ray, acu00157,
Cletus Mitonga, Jawaid Ali 10,024 Registered Users |
Forum Statistics |
Forums25 Topics10,909 Posts73,033 Members10,024 |
Most Online5,980 |
Troubleshooting
Table 7-3
Failure Identification Codes — continued
Code
93
Write data and read data of
M5M5165 do not agree.
94
Configuration option settings
checksum is not 0.
95
Checksum error of alarm log data.
96
Checksum error of ROM code.
99
Undefined failure code received
from Slave CPU.
Table 7-4
Troubleshooting Chart
Symptom
Power Issues
The pump cannot be turned on.
07-19-B1-688
Cause
Turn pump off and back on.
Reset all of the configuration
option settings in accordance with
"Modifying the Configuration
Option Settings" on page 5-2.
Turn pump off and back on.
Check battery voltage in
accordance with "Automatic Test
Mode 1: Calibration Mode 1" on
page 9-14.
Turn pump off and back on. Clear
the alarm log via the modify
configuration mode.
Turn pump off and back on.
Turn pump off and back on.
Table 7-4 identifies specific failures by symptoms rather than Failure
Identification Codes.
Check backup battery voltage in
accordance with "Automatic Test
Mode 1: Calibration Mode 1" on page
9-14.
Check front panel cables for short
circuits and proper connection.
Check CNBUS1/CNBUS2, CN601
and CN602 for proper connection and
continuity.
FLO-GARD 6201 VOLUMETRIC INFUSION PUMP SERVICE MANUAL
Checks
Checks
Troubleshooting Chart
Corrective Action
If failure recurs, replace CPU
board.
If failure recurs, charge or replace
the battery.
Replace CPU board.
If failure recurs, replace CPU
board.
If failure recurs, replace CPU
board.
If failure recurs, replace CPU
board.
Corrective Action
Sometimes this condition is followed
by a F49 and/or F94 code. Charge or
replace the battery to clear the
condition. Repair if necessary.
7-19