Hi,

I need some help. I Have to repair an infusion pump Baxter Flo-Gard 6301 but appears in display COMMON FAILURE F94, someone can help me?

Best regards,


Re: Flo- Gard 6301

Mofe #55222 11/03/11 2:38 AM

Joined: Apr 2009

Posts: 81

hong kong

dennis_yim

Adept

dennis_yim

Adept

Joined: Apr 2009

Posts: 81

hong kong

this is seem a common fault on 6201.
you search a 6201 in discuss forum,
you may be found a answer.

Last edited by dennis_yim; 11/03/11 2:39 AM.

Re: Flo- Gard 6301

Mofe #55226 11/03/11 6:50 AM

Joined: Feb 2004

Posts: 14,483

Likes: 37

the path less trodden

Geoff Hannis

Super Hero

Geoff Hannis

Super Hero

Joined: Feb 2004

Posts: 14,483

Likes: 37

the path less trodden

Check-out this thread, Mofe. We've been there before.


Moderated by  DaveC in Oz, RoJo 

Hop To


Link Copied to Clipboard

Who's Online Now
2 members (Huw, fanie), 104 guests, and 178 robots.

Key: Admin, Global Mod, Mod

Newest Members
JonnyP83, carl ray, acu00157, Cletus Mitonga, Jawaid Ali
10,024 Registered Users

Forum Statistics

Forums25

Topics10,909

Posts73,033

Members10,024

Most Online5,980
Jan 29th, 2020

Troubleshooting

Table 7-3

Failure Identification Codes — continued

Code

93

Write data and read data of

M5M5165 do not agree.

94

Configuration option settings

checksum is not 0.

95

Checksum error of alarm log data.

96

Checksum error of ROM code.

99

Undefined failure code received

from Slave CPU.

Table 7-4

Troubleshooting Chart

Symptom

Power Issues

The pump cannot be turned on.

07-19-B1-688

Cause

Turn pump off and back on.

Reset all of the configuration

option settings in accordance with

"Modifying the Configuration

Option Settings" on page 5-2.

Turn pump off and back on.

Check battery voltage in

accordance with "Automatic Test

Mode 1: Calibration Mode 1" on

page 9-14.

Turn pump off and back on. Clear

the alarm log via the modify

configuration mode.

Turn pump off and back on.

Turn pump off and back on.

Table 7-4 identifies specific failures by symptoms rather than Failure

Identification Codes.

Check backup battery voltage in

accordance with "Automatic Test

Mode 1: Calibration Mode 1" on page

9-14.

Check front panel cables for short

circuits and proper connection.

Check CNBUS1/CNBUS2, CN601

and CN602 for proper connection and

continuity.

FLO-GARD 6201 VOLUMETRIC INFUSION PUMP SERVICE MANUAL

Checks

Checks

Troubleshooting Chart

Corrective Action

If failure recurs, replace CPU

board.

If failure recurs, charge or replace

the battery.

Replace CPU board.

If failure recurs, replace CPU

board.

If failure recurs, replace CPU

board.

If failure recurs, replace CPU

board.

Corrective Action

Sometimes this condition is followed

by a F49 and/or F94 code. Charge or

replace the battery to clear the

condition. Repair if necessary.

7-19